Medical Devices; Hematology and Pathology Devices; Classification of the Digital Cervical Cytology Slide Imaging System With Artificial Intelligence Algorithm
Primary source
Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.
Issuing agencies
Abstract
The Food and Drug Administration (FDA) is classifying the digital cervical cytology slide imaging system with artificial intelligence algorithm into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the digital cervical cytology slide imaging system with artificial intelligence algorithm. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.
Full Text
<html>
<head>
<title>Federal Register, Volume 91 Issue 213 (Thursday, November 5, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 213 (Thursday, November 5, 2026)]
[Rules and Regulations]
[Pages 69687-69690]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-22593]
[[Page 69687]]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 864
[Docket No. FDA-2026-N-12384]
Medical Devices; Hematology and Pathology Devices; Classification
of the Digital Cervical Cytology Slide Imaging System With Artificial
Intelligence Algorithm
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
digital cervical cytology slide imaging system with artificial
intelligence algorithm into class II (special controls). The special
controls that apply to the device type are identified in this order and
will be part of the codified language for classification of the digital
cervical cytology slide imaging system with artificial intelligence
algorithm. We are taking this action because we have determined that
classifying the device into class II will provide a reasonable
assurance of the safety and effectiveness of the device. We believe
this action will also enhance patients' access to beneficial innovative
devices, in part by reducing regulatory burdens.
DATES: This order is effective November 5, 2026. The classification was
applicable on January 31, 2024.
FOR FURTHER INFORMATION CONTACT: Arpita Roy, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3302, Silver Spring, MD 20993-0002, 240-402-4807,
<a href="/cdn-cgi/l/email-protection#347546445d40551a665b4d745250551a5c5c471a535b42"><span class="__cf_email__" data-cfemail="bbfac9cbd2cfda95e9d4c2fbdddfda95d3d3c895dcd4cd">[email protected]</span></a>.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA (the Agency or we) has classified the digital
cervical cytology slide imaging system with artificial intelligence
algorithm into class II (special controls), which we have determined
will provide a reasonable assurance of the safety and effectiveness of
the device. In addition, we believe this action will enhance patients'
access to beneficial innovation, in part by reducing regulatory burdens
by placing the device into a lower device class than the automatic
class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified into, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate device by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo
classification process by adding a second procedure. A device sponsor
may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a premarket
notification (510(k)) for a device that has not previously been
classified. After receiving an order from FDA classifying the device
into class III under section 513(f)(1) of the FD&C Act, the person then
requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device sponsors do not have to submit a
De Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On August 27, 2021, FDA received Hologic Inc.'s request for De Novo
classification of the Genius Digital Diagnostics System with the Genius
Cervical AI algorithm. FDA reviewed the request in order to classify
the device under the criteria for classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of the safety and
effectiveness of the device, but there is sufficient information to
establish special controls that, in combination with the general
controls, provide reasonable assurance of the safety and effectiveness
of the device for its intended use (see section 513(a)(1)(B) of the
FD&C Act). After review of the information submitted in the request, we
determined that the device can be classified into class II with the
establishment of special controls. FDA has determined that these
special controls, in addition to the general controls, will provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, on January 31, 2024, FDA issued an order to the
requester classifying the device into class II. In this final order,
FDA is codifying the classification of the device by adding 21 CFR
864.3900.\1\ We have named the
[[Page 69688]]
generic type of device ``digital cervical cytology slide imaging system
with artificial intelligence algorithm,'' and it is identified as a
prescription in vitro diagnostic device consisting of an automated
digital slide creation, viewing, and management system intended to aid
in the review of digital images of slides prepared from Pap test
specimens and conventional Pap smears by selecting and presenting areas
of interest to facilitate interpretation by the reader. The artificial
intelligence (AI) software algorithm uses machine learning techniques
to provide information to the user about the presence, location, or
characteristics of areas of the image with clinical implications.
Information from this device is intended to assist the user in
rendering a cytology diagnosis.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the risks to health associated with this type of
device and the measures required to mitigate these risks in table 1.
Table 1--Risks to Health and Mitigation Measures for Digital Cervical
Cytology Slide Imaging Systems With Artificial Intelligence Algorithm
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
False negative and false positive Certain design verification and
results due to device errors in validation, including certain
presenting abnormal cells in the studies and risk mitigation
digital image gallery. analysis.
Certain labeling information,
including limitations, device
descriptions, methodology and
protocols, and performance
information.
False negative and false positive Certain design verification and
results due to incorrect validation, including certain
interpretation of digital images or studies and risk mitigation
incorrect operation of the device by analysis.
the user. Certain labeling information,
including limitations, user
training, device descriptions,
methodology and protocols, and
performance information.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of the safety and effectiveness of the device. For a device
to fall within this classification, and thus avoid automatic
classification in class III, it would have to comply with the special
controls named in this final order. The necessary special controls
appear in the regulation codified by this final order.
At the time of classification, digital cervical cytology slide
imaging systems with AI algorithm are for prescription use only.
Therefore, these devices are subject to the prescription labeling
requirements for in vitro diagnostic products (see 21 CFR 809.10(a)(4)
and (b)(5)(ii)).
Under the FD&C Act, submission of a premarket notification under
section 510(k) is required to reasonably assure the safety and
effectiveness of class II devices unless FDA determines that the device
type should be exempt under section 510(m) of the FD&C Act. At this
time, FDA has not made this determination for digital cervical cytology
slide imaging systems with AI algorithm. This device is therefore
subject to premarket notification requirements under section 510(k) of
the FD&C Act.
III. Analysis of Environmental Impact
We have determined under 21 CFR 25.34(b) that this action is of a
type that does not normally have a significant effect on the human
environment. Therefore, neither an environmental assessment nor an
environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions have been approved under OMB control
number 0910-0120; the collections of information in 21 CFR part 820
regarding quality management system regulation have been approved under
OMB control number 0910-0073; and the collections of information in 21
CFR parts 801 and 809 regarding labeling have been approved under OMB
control number 0910-0485.
List of Subjects in 21 CFR Part 864
Blood, Medical devices, Packaging and containers.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
864 is amended as follows:
PART 864--HEMATOLOGY AND PATHOLOGY DEVICES
0
1. The authority citation for part 864 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 864.3900 to subpart D to read as follows:
Sec. 864.3900 Digital cervical cytology slide imaging system with
artificial intelligence algorithm.
(a) Identification. The digital cervical cytology slide imaging
system with artificial intelligence (AI) algorithm is a prescription in
vitro diagnostic device consisting of an automated digital slide
creation, viewing, and management system intended to aid in the review
of digital images of slides prepared from Pap test specimens and
conventional Pap smears by selecting and presenting areas of interest
to facilitate interpretation by the reader. The AI software algorithm
uses machine learning techniques to provide information to the user
about the presence, location, or characteristics of areas of the image
with clinical implications. Information from this device is intended to
assist the user in rendering a cytology diagnosis.
(b) Classification. Class II (special controls). The special
controls for this device are:
[[Page 69689]]
(1) The labeling required under Sec. 809.10(b) of this chapter
must include the following:
(i) The intended use must include the specimen type (e.g., Pap test
or conventional Pap smear), information on the device input (e.g., Pap
test glass slides) and output(s) (e.g., whole slide image (WSI) of
cytology glass slides or cell spot), information on the AI algorithm
output(s) (i.e., format of the information provided by the device to
the user, such as objects of interest (OOI) found on the WSI, which is
presented in a gallery for the user to review), and intended users;
(ii) A detailed description of the intended patient population;
(iii) Detailed instructions for use, including the intended image
review and interpreting protocol for use of the device, and how to
resolve device-related issues (e.g., cybersecurity or device
malfunction issues);
(iv) A detailed description of the device, including the following:
(A) Necessary input/output devices (e.g., scanner, imager,
management server, review station, viewing software, software
algorithm);
(B) Necessary infrastructure, platform, and/or software application
services, and configuration(s) (e.g., storage, network architecture,
cloud services);
(C) Compatible scanning hardware and scanning protocols;
(D) Interaction with input/output devices, and necessary third-
party software;
(E) Recommended training for intended users for safe use of the
device;
(v) A detailed summary of the performance testing required by
paragraphs (b)(2)(i), (b)(2)(ii), (b)(2)(iv), and (b)(2)(v)(A) of this
section, including test methods, dataset characteristics, results, and
a summary of sub-analyses on case distributions stratified by relevant
confounders, i.e., three-dimensional (3D) characteristics of cells and
cell clusters in cytology samples, patient demographics, medical
history, user experience, sites, and scanning equipment as applicable;
(vi) Intended user screening time data to provide information about
daily slide screening workload limits, as determined by the performance
testing in paragraph (b)(2)(v)(B); and
(vii) Limiting statements, including the following:
(A) A description of situations in which the device may fail or may
not operate at its expected performance level (e.g., poor image quality
or for certain subpopulations), including any limitations due to the
dataset used to train, test, and tune the algorithm during device
development;
(B) That review of the entire gallery (i.e., OOIs with clinical
implications) should be performed by the user prior to rendering a
diagnosis;
(C) That users should use the device in conjunction with the
standard of care, an assessment of patient history and other risk
factors, and professional screening and management guidelines to guide
patient care; and
(D) That users should employ appropriate procedures and safeguards
(e.g., daily quality control measures, deferral to glass slide review,
cybersecurity), including when the device is used for remote review of
images, to assure the validity of the interpretation of images obtained
using this device.
(2) Design verification and validation must include:
(i) A detailed description and the technical performance assessment
of each component in the device, including the following, as
appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Z-focal plane acquisition, including a detailed description of
methods used to ensure that the complete depth of the cellular material
is captured (e.g., Z-plane merging, Z-stacking), when applicable;
(E) Mechanical scanner movement;
(F) Digital imaging sensor;
(G) Image processing software;
(H) Image composition techniques;
(I) Image file formats;
(J) Image review manipulation software;
(K) Computer environment; and
(L) Display system.
(ii) Detailed technical performance assessment at the system level,
including for the following, as appropriate:
(A) Coverage of Z-depth to represent the 3D aspects of cells and
cell clusters in cytology samples;
(B) Color reproducibility;
(C) Spatial resolution;
(D) Focusing accuracy;
(E) Whole slide cell spot or specimen material coverage;
(F) Stitching error; and
(G) Turnaround time.
(iii) A detailed description of the device software, including its
algorithm and its development, that includes a description of any
datasets used to train, tune, or test the software algorithm. This
detailed description of the device software must include:
(A) The study protocols (e.g., OOI reproducibility study, imaging
reproducibility study, cell count validation study) and results used to
assess the device output(s) (e.g., image overlays, image heatmaps,
image location, total number of cells);
(B) The training dataset, including cases representing different
pre-analytical variables representative of the conditions likely to be
encountered when used as intended (e.g., fixation type and time,
cytology slide processors, challenging diagnostic cases, multiple
sites, patient demographics);
(C) An appropriate number of digital cytology images in an
independent validation dataset to demonstrate device accuracy in
detecting OOI on scanned cytology images, which must include subsets
clinically relevant to the intended use of the device; and
(D) A system level architecture diagram with a matrix to depict the
communication endpoints, communication protocols, and security
protections for the device and its supportive systems, including any
products or services that are included in the communication pathway.
(iv) Detailed documentation of studies demonstrating acceptable, as
determined by FDA, analytical device performance using samples that are
representative of the entire spectrum of challenging cases likely to be
encountered when the device is used as intended. For each analytical
study, relevant details must be documented (e.g., the origin of the
study slides and images, reader/annotator qualifications, method of
annotation, location of the study site(s), challenging diagnoses). The
analytical studies must include:
(A) Testing to assess device outputs (e.g., OOI reproducibility
study, cell count validation study); and
(B) Precision and reproducibility study(ies) that demonstrate
device performance when used with multiple input devices (e.g., digital
cytology scanners) to assess total variability across images,
operators, within-scanners, between-scanners and between-sites, using a
comprehensive set of representative clinical specimens with defined,
clinically relevant features and challenging characteristics likely to
be encountered when the device is used as intended. Precision and
reproducibility must be assessed by agreement between replicates.
(v) Data demonstrating acceptable, as determined by FDA, clinical
validation by conducting studies with clinical specimens to demonstrate
that screening and diagnosing digital images of cervical cytology
slides with the assistance of the device's software algorithms is non-
inferior to using a comparator method that FDA has determined to be
appropriate (e.g., optical microscope). For each clinical study,
relevant details must be
[[Page 69690]]
documented (e.g., the origin of the study slides and images, reader/
annotator qualifications, method of annotation, location of the study
site(s) (on-site/remote), challenging diagnoses). The studies must
include:
(A) A study demonstrating the performance by the intended users
with and without the AI algorithm (i.e., unassisted (manual microscope
read) and device-assisted reading of scanned digital images of cervical
cytology slides). The study dataset must contain sufficient numbers of
cytological slides for each diagnostic category and processed by
representative processing instruments such that the performance
estimates and confidence intervals for these individual subsets can be
appropriately characterized. The performance assessment must be based
on appropriate diagnostic accuracy measures (e.g., sensitivity,
specificity, predictive value, diagnostic likelihood ratio); and
(B) Intended user screening time data obtained from the clinical
study, to characterize and set recommended daily maximum workload
screening limits for intended users. The study dataset must contain
sufficient cases from relevant cohorts that are representative of the
scope of patients likely to be encountered given the intended use of
the device (e.g., subsets defined by clinically relevant confounders,
challenging diagnoses, subsets with potential biopsy appearance
modifiers, concomitant diseases, subsets defined by image scanning
characteristics). The reader population must be comprised of the
intended user population in terms of clinical training, certification,
and years of experience.
(vi) Human factors study data to evaluate the usability and user
experience of the device with representative users.
Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-22593 Filed 11-4-26; 8:45 am]
BILLING CODE 4164-01-P
</pre><script data-cfasync="false" src="/cdn-cgi/scripts/5c5dd728/cloudflare-static/email-decode.min.js"></script></body>
</html>This is legal information, not legal advice. Laws vary by jurisdiction and change frequently. Always verify current law with official sources and consult a licensed attorney in your jurisdiction for advice on your specific situation.