Medical Devices; Neurological Devices; Classification of the Prognostic Assessment Software of Mild Cognitive Impairment
Primary source
Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.
Issuing agencies
Abstract
The Food and Drug Administration (FDA) is classifying the prognostic assessment software of mild cognitive impairment into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the prognostic assessment software of mild cognitive impairment. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.
Full Text
<html>
<head>
<title>Federal Register, Volume 91 Issue 213 (Thursday, November 5, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 213 (Thursday, November 5, 2026)]
[Rules and Regulations]
[Pages 69690-69692]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-22590]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 882
[Docket No. FDA-2026-N-12212]
Medical Devices; Neurological Devices; Classification of the
Prognostic Assessment Software of Mild Cognitive Impairment
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
prognostic assessment software of mild cognitive impairment into class
II (special controls). The special controls that apply to the device
type are identified in this order and will be part of the codified
language for classification of the prognostic assessment software of
mild cognitive impairment. We are taking this action because we have
determined that classifying the device into class II will provide a
reasonable assurance of the safety and effectiveness of the device. We
believe this action will also enhance patients' access to beneficial
innovative devices, in part by reducing regulatory burdens.
DATES: This order is effective November 5, 2026. The classification was
applicable on January 11, 2024.
FOR FURTHER INFORMATION CONTACT: Patrick Antkowiak, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 4254, Silver Spring, MD 20993-0002, 240-
402-3705, <a href="/cdn-cgi/l/email-protection#500031242239333b7e113e243b3f2739313b103634317e3838237e373f26"><span class="__cf_email__" data-cfemail="f5a59481879c969edbb49b819e9a829c949eb5939194db9d9d86db929a83">[email protected]</span></a>.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA (the Agency or we) has classified the prognostic
assessment software of mild cognitive impairment into class II (special
controls), which we have determined will provide a reasonable assurance
of the safety and effectiveness of the device. In addition, we believe
this action will enhance patients' access to beneficial innovation, in
part by reducing regulatory burdens by placing the device into a lower
device class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified into, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate device by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo
classification process by adding a second procedure. A device sponsor
may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a premarket
notification (510(k)) for a device that has not previously been
classified. After receiving an order from FDA classifying the device
into class III under section 513(f)(1) of the FD&C Act, the person then
requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically
[[Page 69691]]
placed within class III, the De Novo classification is considered to be
the initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device sponsors do not have to submit a
De Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On September 29, 2022, FDA received Darmiyan, Inc.'s request for De
Novo classification of the BrainSee device. FDA reviewed the request in
order to classify the device under the criteria for classification set
forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of the safety and
effectiveness of the device, but there is sufficient information to
establish special controls that, in combination with the general
controls, provide reasonable assurance of the safety and effectiveness
of the device for its intended use (see section 513(a)(1)(B) of the
FD&C Act). After review of the information submitted in the request, we
determined that the device can be classified into class II with the
establishment of special controls. FDA has determined that these
special controls, in addition to the general controls, will provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, on January 11, 2024, FDA issued an order to the
requester classifying the device into class II. In this final order,
FDA is codifying the classification of the device by adding 21 CFR
882.1454.\1\ We have named the generic type of device ``prognostic
assessment software of mild cognitive impairment,'' and it is
identified as a prescription-only device that combines information from
multiple clinical data sources to produce an output indicative of the
likelihood that a patient diagnosed with mild cognitive impairment will
progress to more severe forms of cognitive impairment, or will develop
dementia, such as Alzheimer's disease dementia, within a specified
period of time. This device is intended for adjunctive use and not
intended as a stand-alone diagnostic or prognostic tool.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the risks to health associated with this type of
device and the measures required to mitigate these risks in table 1.
Table 1--Risks to Health and Mitigation Measures for Prognostic
Assessment Software of Mild Cognitive Impairment
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Device failure or incorrect analysis Clinical performance testing;
leading to Postmarket surveillance;
<bullet> False positives or falsely Software verification,
high prognostic outputs resulting in validation, and hazard
inappropriate patient treatment and analysis; and Labeling.
delayed diagnosis of non-Alzheimer's
disease conditions..
<bullet> False negatives or falsely low
prognostic outputs resulting in
delayed diagnosis and/or patient
treatment..
Use error or misinterpretation of Human factors/usability
results leading to. assessment; and Labeling.
<bullet> A false positive or
falsely high prognostic outputs.
<bullet> A false negative or
falsely low prognostic outputs.
<bullet> Overreliance on device
output for follow-up.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of the safety and effectiveness of the device. For a device
to fall within this classification, and thus avoid automatic
classification in class III, it would have to comply with the special
controls named in this final order. The necessary special controls
appear in the regulation codified by this final order.
At the time of classification, prognostic assessment software of
mild cognitive impairment is for prescription use only. Prescription
devices are exempt from the requirement for adequate directions for use
for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR
801.109 are met.
Under the FD&C Act, submission of a premarket notification under
section 510(k) is required to reasonably assure the safety and
effectiveness of class II devices unless FDA determines that the device
type should be exempt under section 510(m) of the FD&C Act. At this
time FDA has not made this determination for prognostic assessment
software of mild cognitive impairment. This device is therefore subject
to premarket notification requirements under section 510(k) of the FD&C
Act.
III. Analysis of Environmental Impact
We have determined under 21 CFR 25.34(b) that this action is of a
type that does not normally have a significant effect on the human
environment. Therefore, neither an environmental assessment nor an
environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval have been
[[Page 69692]]
approved under OMB control number 0910-0231; the collections of
information in part 807, subpart E, regarding premarket notification
submissions have been approved under OMB control number 0910-0120; the
collections of information in 21 CFR part 820 regarding quality
management system regulation have been approved under OMB control
number 0910-0073; and the collections of information in 21 CFR part 801
regarding labeling have been approved under OMB control number 0910-
0485.
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
882 is amended as follows:
PART 882--NEUROLOGICAL DEVICES
0
1. The authority citation for part 882 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 882.1454 to subpart B to read as follows:
Sec. 882.1454 Prognostic assessment software of mild cognitive
impairment.
(a) Identification. Prognostic assessment software of mild
cognitive impairment is a prescription-only device that combines
information from multiple clinical data sources to produce an output
indicative of the likelihood that a patient diagnosed with mild
cognitive impairment will progress to more severe forms of cognitive
impairment, or will develop dementia, such as Alzheimer's disease
dementia, within a specified period of time. This device is intended
for adjunctive use and not intended as a stand-alone diagnostic or
prognostic tool.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Data obtained from premarket clinical performance validation
testing and postmarket surveillance acquired under anticipated
conditions of use must demonstrate that the device performs as intended
when used to analyze data from the intended patient population, unless
FDA determines, based on the totality of the information provided for
premarket review, that data from postmarket surveillance is not
required.
(i) Data provided from paragraph (b)(1) of this section must
demonstrate the accuracy, precision, and reproducibility of the device
for predicting progression of patients with mild cognitive impairment
to more severe forms of cognitive impairment or to dementia, such as
Alzheimer's disease dementia, based on a clinically relevant reference
standard for the diagnosis of such disease states or conditions. The
test data set must include data acquired from a patient population that
is representative of the intended patient population normally seen in
clinical practice. The test data set must be independent from data used
in training/development of the device.
(ii) Objective performance measures (e.g., sensitivity,
specificity, positive predictive value, negative predictive value) must
be reported with relevant descriptive or developmental performance
measures.
(2) Software verification, validation, and hazard analysis must be
provided. Software documentation must include a detailed, technical
description of the model/algorithm(s) and algorithm inputs and outputs.
(3) Human factors/usability assessment must demonstrate that the
intended user(s) in the intended use environment can correctly use the
device and interpret the device output(s).
(4) Physician labeling must include:
(i) A detailed summary of the clinical performance testing methods,
including results of the performance testing for tested performance
measures/metrics, selection criteria, and the patient demographics;
(ii) A description of the patient population that was used in
development or training of the device algorithm/model;
(iii) Device limitations or subpopulations for which the device may
not perform as expected or for whom the device has not been validated;
(iv) A statement that the device is not a stand-alone diagnostic or
prognostic tool and that the device output should only be interpreted
in the context of all available clinical information, including patient
history, medications, and comorbidities; and
(v) Information for interpretation of the device outputs detailing
the risks associated with misinterpretation of the device outputs.
(5) Patient labeling must include information explaining the device
output(s) and the associated risks if the device output is incorrect.
Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-22590 Filed 11-4-26; 8:45 am]
BILLING CODE 4164-01-P