Medical Devices; General and Plastic Surgery Devices; Classification of the Software-Aided Adjunctive Diagnostic Device for Use by Physicians on Lesions Suspicious for Skin Cancer
Primary source
Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.
Issuing agencies
Abstract
The Food and Drug Administration (FDA) is classifying the software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.
Full Text
<html>
<head>
<title>Federal Register, Volume 91 Issue 209 (Friday, October 30, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 209 (Friday, October 30, 2026)]
[Rules and Regulations]
[Pages 68465-68467]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-22179]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 878
[Docket No. FDA-2026-N-12213]
Medical Devices; General and Plastic Surgery Devices;
Classification of the Software-Aided Adjunctive Diagnostic Device for
Use by Physicians on Lesions Suspicious for Skin Cancer
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
software-aided adjunctive diagnostic device for use by physicians on
lesions suspicious for skin cancer into class II (special controls).
The special controls that apply to the device type are identified in
this order and will be part of the codified language for classification
of the software-aided adjunctive diagnostic device for use by
physicians on lesions suspicious for skin cancer. We are taking this
action because we have determined that classifying the device into
class II will provide a reasonable assurance of the safety and
effectiveness of the device. We believe this action will also enhance
patients' access to beneficial innovative devices, in part by reducing
regulatory burdens.
DATES: This order is effective October 30, 2026. The classification was
applicable on January 12, 2024.
FOR FURTHER INFORMATION CONTACT: Rudy Andriani, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4540, Silver Spring, MD 20993-0002, 240-402-6030,
<a href="/cdn-cgi/l/email-protection#df8daabba6f19eb1bbadb6beb1b69fb9bbbef1b7b7acf1b8b0a9"><span class="__cf_email__" data-cfemail="326047564b1c735c56405b535c5b725456531c5a5a411c555d44">[email protected]</span></a>.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA (the Agency or we) has classified the software-
aided adjunctive diagnostic device for use by physicians on lesions
suspicious for skin cancer into class II (special controls), which we
have determined will provide a reasonable assurance of the safety and
effectiveness of the device. In addition, we believe this action will
enhance patients' access to beneficial innovation, in part by reducing
regulatory burdens by placing the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified into, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate device by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo
classification process by adding a second procedure. A device sponsor
may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a premarket
notification (510(k)) for a device that has not previously been
classified. After receiving an order from FDA classifying the device
into class III under section 513(f)(1) of the FD&C Act, the person then
requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device sponsors do not have to submit a
De Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On February 2, 2023, FDA received DermaSensor Inc.'s request for De
Novo classification of the DermaSensor device. FDA reviewed the request
in order to classify the device under the criteria for classification
set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable
[[Page 68466]]
assurance of the safety and effectiveness of the device, but there is
sufficient information to establish special controls that, in
combination with the general controls, provide reasonable assurance of
the safety and effectiveness of the device for its intended use (see
section 513(a)(1)(B) of the FD&C Act). After review of the information
submitted in the request, we determined that the device can be
classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on January 12, 2024, FDA issued an order to the
requester classifying the device into class II. In this final order,
FDA is codifying the classification of the device by adding 21 CFR
878.1830.\1\ We have named the generic type of device ``software-aided
adjunctive diagnostic device for use by physicians on lesions
suspicious for skin cancer,'' and it is identified as a prescription
device that uses a software algorithm to analyze optical or other
physical properties of a skin lesion and returns a classification of
the skin lesion. The device is intended for use by a physician not
trained in the clinical diagnosis and management of skin cancer as an
adjunctive second-read device following identification of a suspicious
skin lesion. It is not for use as a standalone diagnostic and is not
for use to confirm a clinical diagnosis.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the risks to health associated with this type of
device and the measures required to mitigate these risks in table 1.
Table 1--Risks to Health and Mitigation Measures for Software-Aided
Adjunctive Diagnostic Device for Use by Physicians on Lesions Suspicious
for Skin Cancer
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
False negative results, leading to Clinical performance testing;
failure to treat cancer and cancer Postmarket surveillance; Non-
progression, or false positive clinical performance testing;
results, leading to unnecessary and Labeling.
referrals and/or medical procedures.
False results or failure to generate a Precision testing; Human
result due to use error or improper factors testing; and Labeling.
device use.
False results or failure to generate a Non-clinical performance
result due to device failure or testing; Precision testing;
malfunction. Software verification,
validation, and hazard
analysis; and Labeling.
Electrical, thermal, mechanical, or Electrical, mechanical, and
light exposure-related injury. thermal safety testing;
Software verification,
validation, and hazard
analysis; and Labeling.
Interference with other devices........ Electromagnetic compatibility
testing.
Adverse tissue reaction................ Biocompatibility evaluation.
Infection and cross-contamination...... Cleaning and disinfection
validation; and Labeling.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of the safety and effectiveness of the device. For a device
to fall within this classification, and thus avoid automatic
classification in class III, it would have to comply with the special
controls named in this final order. The necessary special controls
appear in the regulation codified by this final order.
At the time of classification, software-aided adjunctive diagnostic
devices for use by physicians on lesions suspicious for skin cancer are
for prescription use only. Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR
801.5, as long as the conditions of 21 CFR 801.109 are met.
Under the FD&C Act, submission of a premarket notification under
section 510(k) is required to reasonably assure the safety and
effectiveness of class II devices unless FDA determines that the device
type should be exempt under section 510(m) of the FD&C Act. At this
time FDA has not made this determination for software-aided adjunctive
diagnostic devices for use by physicians on lesions suspicious for skin
cancer. This device is therefore subject to premarket notification
requirements under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
We have determined under 21 CFR 25.34(b) that this action is of a
type that does not normally have a significant effect on the human
environment. Therefore, neither an environmental assessment nor an
environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions have been approved under OMB control
number 0910-0120; the collections of information in 21 CFR part 820
regarding quality management system regulation have been approved under
OMB control number 0910-0073; and the collections of information in 21
CFR part 801 regarding labeling have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 878
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
878 is amended as follows:
[[Page 68467]]
PART 878--GENERAL AND PLASTIC SURGERY DEVICES
0
1. The authority citation for part 878 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 878.1830 to subpart B to read as follows:
Sec. 878.1830 Software-aided adjunctive diagnostic device for use by
physicians on lesions suspicious for skin cancer.
(a) Identification. A software-aided adjunctive diagnostic device
for use by physicians on lesions suspicious for skin cancer is a
prescription device that uses a software algorithm to analyze optical
or other physical properties of a skin lesion and returns a
classification of the skin lesion. The device is intended for use by a
physician not trained in the clinical diagnosis and management of skin
cancer as an adjunctive second-read device following identification of
a suspicious skin lesion. It is not for use as a standalone diagnostic
and is not for use to confirm a clinical diagnosis.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Data obtained from premarket clinical performance validation
testing and postmarket surveillance acquired under anticipated
conditions of use must demonstrate that the device performs as intended
in the intended patient population, unless FDA determines based on the
totality of the information provided for premarket review that data
from postmarket surveillance is not required.
(i) Data must demonstrate superior accuracy of device-aided users'
diagnostic characterization of the indicated lesions compared to the
accuracy of unaided users.
(ii) Clinical testing must evaluate patients across a range of skin
phototypes, risk factors, and anatomic areas that represents the
intended use population.
(iii) Standalone device performance testing must demonstrate the
accuracy of the device output relative to ground truth, including the
following:
(A) Testing must demonstrate at least 90 percent sensitivity of the
device output for lesions with high metastatic potential, or an
alternative clinical consideration must be provided to justify lower
sensitivity. Clinical justification must be provided for the reported
specificity.
(B) Lesions must be selected by representative users and include a
justified quantity and range of mimic lesions per diagnosis.
(C) Justification must be provided for the determination of ground
truth.
(D) Testing must include a representative range of individuals with
diverse risk factors (including age, body site, skin phototype, and
other clinical factors), and analysis of standalone performance must
include subgroup analysis by relevant risk factors.
(2) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use,
including compatibility testing of the device software with specific
signal or image acquisition hardware. Testing must include a
description of compatible hardware and processes, pre-specified
compatibility testing protocols, and dataset(s).
(3) Performance testing must demonstrate device precision,
including repeatability and reproducibility of device performance,
across operators and challenging use conditions.
(4) Performance testing must demonstrate the electrical safety,
mechanical safety, thermal safety, and electromagnetic compatibility of
any electrical components of the device.
(5) Performance testing must validate reprocessing instructions for
reusable components of the device.
(6) The patient-contacting components of the device must be
demonstrated to be biocompatible.
(7) Software verification, validation, and hazard analysis must be
performed.
(8) Human factors assessment must demonstrate that the device can
be safely and correctly used by intended users.
(9) Labeling must include:
(i) A summary of standalone and clinical performance testing
conducted with the device that describes performance measures,
including sensitivity and specificity, and statistical confidence
intervals, as well as performance of the device for all clinically
relevant subgroups within the intended use population;
(ii) A description of the patient population that was used in
development or training of the device algorithm;
(iii) Device limitations or subpopulations for which the device may
not perform as expected or for whom the device has not been validated;
(iv) Information for interpretation of the device outputs detailing
the risks associated with misinterpretation of the device outputs;
(v) Warnings to avoid unsafe exposure to any energy-emitting
components of the device, e.g., excluding lesions close to the eye;
(vi) A statement that the device is not intended for use as a
standalone diagnostic; and
(vii) Instructions for device maintenance and validated methods and
instructions for reprocessing of any reusable components.
Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-22179 Filed 10-29-26; 8:45 am]