Medical Devices; General and Plastic Surgery Devices; Classification of the Focused Ultrasound System for Non-Thermal, Mechanical Tissue Ablation
Primary source
Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.
Issuing agencies
Abstract
The Food and Drug Administration (FDA) is classifying the focused ultrasound system for non-thermal, mechanical tissue ablation into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the focused ultrasound system for non-thermal, mechanical tissue ablation. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.
Full Text
<html>
<head>
<title>Federal Register, Volume 91 Issue 192 (Tuesday, October 6, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 192 (Tuesday, October 6, 2026)]
[Rules and Regulations]
[Pages 63493-63495]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-20440]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 878
[Docket No. FDA-2026-N-10992]
Medical Devices; General and Plastic Surgery Devices;
Classification of the Focused Ultrasound System for Non-Thermal,
Mechanical Tissue Ablation
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
focused ultrasound system for non-thermal, mechanical tissue ablation
into class II (special controls). The special controls that apply to
the device type are identified in this order and will be part of the
codified language for classification of the focused ultrasound system
for non-thermal, mechanical tissue ablation. We are taking this action
because we have determined that classifying the device into class II
will provide a reasonable assurance of the safety and effectiveness of
the device. We believe this action will also enhance patients' access
to beneficial innovative devices, in part by reducing regulatory
burdens.
DATES: This order is effective October 6, 2026. The classification was
applicable on October 6, 2023.
FOR FURTHER INFORMATION CONTACT: Cal Rabang, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4633, Silver Spring, MD 20993-0002, 301-796-6412,
<a href="/cdn-cgi/l/email-protection#b2f1d3de9ce0d3d0d3dcd5f2d4d6d39cdadac19cd5ddc4"><span class="__cf_email__" data-cfemail="93d0f2ffbdc1f2f1f2fdf4d3f5f7f2bdfbfbe0bdf4fce5">[email protected]</span></a>.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA (the Agency or we) has classified the focused
ultrasound system for non-thermal, mechanical tissue ablation into
class II (special controls), which we have determined will provide a
reasonable assurance of the safety and effectiveness of the device. In
addition, we believe this action will enhance patients' access to
beneficial innovation, in part by
[[Page 63494]]
reducing regulatory burdens by placing the device into a lower device
class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified into, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate device by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo
classification process by adding a second procedure. A device sponsor
may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a premarket
notification (510(k)) for a device that has not previously been
classified. After receiving an order from FDA classifying the device
into class III under section 513(f)(1) of the FD&C Act, the person then
requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device sponsors do not have to submit a
De Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On December 2, 2022, FDA received HistoSonics, Inc.'s request for
De Novo classification of the Edison System. FDA reviewed the request
in order to classify the device under the criteria for classification
set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of the safety and
effectiveness of the device, but there is sufficient information to
establish special controls that, in combination with the general
controls, provide reasonable assurance of the safety and effectiveness
of the device for its intended use (see section 513(a)(1)(B) of the
FD&C Act). After review of the information submitted in the request, we
determined that the device can be classified into class II with the
establishment of special controls. FDA has determined that these
special controls, in addition to the general controls, will provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, on October 6, 2023, FDA issued an order to the requester
classifying the device into class II. In this final order, FDA is
codifying the classification of the device by adding 21 CFR
878.4405.\1\ We have named the generic type of device ``focused
ultrasound system for non-thermal, mechanical tissue ablation,'' and it
is identified as a device that uses focused ultrasound to mechanically
ablate soft tissue. The device is not intended to thermally ablate
tissue.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the risks to health associated with this type of
device and the measures required to mitigate these risks in table 1.
Table 1--Risks to Health and Mitigation Measures for Focused Ultrasound
System for Non-Thermal, Mechanical Tissue Ablation
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Impaired tissue or organ function, Clinical performance testing;
abscess, pain, or other adverse events Labeling; and Human factors
downstream of tissue ablation. testing.
Acoustic path, non-targeted tissue Clinical performance testing;
injury. Human factors testing; Animal
performance testing; and Non-
clinical performance testing.
Tissue injury due to device malfunction Clinical performance testing;
or misuse. Software verification,
validation, and hazard
analysis; Non-clinical
performance testing; Labeling;
and Human factors testing.
Adverse tissue reaction................ Biocompatibility evaluation.
Electrical shock or electromagnetic Electrical safety testing;
interference. Electromagnetic compatibility
testing; and Labeling.
------------------------------------------------------------------------
[[Page 63495]]
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of the safety and effectiveness of the device. For a device
to fall within this classification, and thus avoid automatic
classification in class III, it would have to comply with the special
controls named in this final order. The necessary special controls
appear in the regulation codified by this final order. FDA supports the
principles of the ``3Rs,'' to replace, reduce, and/or refine animal use
in testing when feasible. We encourage sponsors to consult with us if
they wish to use a non-animal testing method they believe is suitable,
adequate, validated, and feasible. We will consider whether such an
alternative method could be assessed for equivalency to an animal test
method.
Under the FD&C Act, submission of a premarket notification under
section 510(k) is required to reasonably assure the safety and
effectiveness of class II devices unless FDA determines that the device
type should be exempt under section 510(m) of the FD&C Act. At this
time FDA has not made this determination for focused ultrasound systems
for non-thermal, mechanical tissue ablation. This device is therefore
subject to premarket notification requirements under section 510(k) of
the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not normally have a significant effect on the human
environment. Therefore, neither an environmental assessment nor an
environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions have been approved under OMB control
number 0910-0120; the collections of information in 21 CFR part 820
regarding quality management system regulation have been approved under
OMB control number 0910-0073; and the collections of information in 21
CFR part 801 regarding labeling have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 878
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
878 is amended as follows:
PART 878--GENERAL AND PLASTIC SURGERY DEVICES
0
1. The authority citation for part 878 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 878.4405 to subpart E to read as follows:
Sec. 878.4405 Focused ultrasound system for non-thermal, mechanical
tissue ablation.
(a) Identification. This device uses focused ultrasound to
mechanically ablate soft tissue. The device is not intended to
thermally ablate tissue.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Clinical performance testing must demonstrate that the device
performs as intended under anticipated conditions of use. Testing must
document the adverse event profile and characterize tissue destruction.
(2) Animal performance testing must demonstrate that the device
mechanically destroys targeted tissue while characterizing the chronic
safety profile, including thermal and mechanical injury to adjacent,
non-target tissue.
(3) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance characteristics must be tested:
(i) Acoustic characterization at clinically relevant settings;
(ii) Determination of the minimum drive voltages necessary to
sustain a bubble cloud in the target tissue;
(iii) Validation of mechanisms to prevent energy delivery that
would result in adverse thermal effects;
(iv) Availability of real-time monitoring during the procedure;
(v) Validation that the treatment zone is limited to the defined
target tissue; and
(vi) Validation of mechanisms to prevent, pause, and terminate
ablation in the event of device failure.
(4) Performance data must support the electrical safety and
electromagnetic compatibility of the device.
(5) All patient-contacting components of the device must be
demonstrated to be biocompatible.
(6) Software validation, verification, and hazard analysis must be
performed.
(7) Human factors testing must demonstrate that the user can safely
and correctly use the device.
(8) Labeling must include the following:
(i) A warning that focused ultrasound ablation should only be
considered in patients with sufficient functional reserve to withstand
the destruction of the planned volume of tissue;
(ii) A statement that the device has not been evaluated for the
treatment of any specific disease or condition; and
(iii) A detailed summary of the clinical testing with the device.
Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-20440 Filed 10-5-26; 8:45 am]
BILLING CODE 4164-01-P
</pre><script data-cfasync="false" src="/cdn-cgi/scripts/5c5dd728/cloudflare-static/email-decode.min.js"></script></body>
</html>This is legal information, not legal advice. Laws vary by jurisdiction and change frequently. Always verify current law with official sources and consult a licensed attorney in your jurisdiction for advice on your specific situation.