Medical Devices; Cardiovascular Devices; Classification of the Mechanical Deviation Device for Esophageal Protection During Cardiac Ablation Procedures

Primary source

Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.

Published
September 10, 2026
Effective
September 10, 2026

Issuing agencies

Health and Human Services DepartmentFood and Drug Administration

Abstract

The Food and Drug Administration (FDA) is classifying the mechanical deviation device for esophageal protection during cardiac ablation procedures into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the mechanical deviation device for esophageal protection during cardiac ablation procedures. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

Full Text

<html>
<head>
<title>Federal Register, Volume 91 Issue 174 (Thursday, September 10, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 174 (Thursday, September 10, 2026)]
[Rules and Regulations]
[Pages 57495-57497]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-18433]


-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 870

[Docket No. FDA-2026-N-9905]


Medical Devices; Cardiovascular Devices; Classification of the 
Mechanical Deviation Device for Esophageal Protection During Cardiac 
Ablation Procedures

AGENCY: Food and Drug Administration, HHS.

ACTION: Final amendment; final order.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA) is classifying the 
mechanical deviation device for esophageal protection during cardiac

[[Page 57496]]

ablation procedures into class II (special controls). The special 
controls that apply to the device type are identified in this order and 
will be part of the codified language for classification of the 
mechanical deviation device for esophageal protection during cardiac 
ablation procedures. We are taking this action because we have 
determined that classifying the device into class II will provide a 
reasonable assurance of the safety and effectiveness of the device. We 
believe this action will also enhance patients' access to beneficial 
innovative devices, in part by reducing regulatory burdens.

DATES: This order is effective September 10, 2026. The classification 
was applicable on September 6, 2023.

FOR FURTHER INFORMATION CONTACT: Aneesh Deoras, Center for Devices and 
Radiological Health, Food and Drug Administration, 10903 New Hampshire 
Ave., Bldg. 66, Rm. 5540, Silver Spring, MD 20993-0002, 240-402-4363, 
<a href="/cdn-cgi/l/email-protection#b4f5dad1d1c7dc9af0d1dbc6d5c7f4d2d0d59adcdcc79ad3dbc2"><span class="__cf_email__" data-cfemail="2a6b444f4f5942046e4f45584b596a4c4e4b04424259044d455c">[email&#160;protected]</span></a>.

SUPPLEMENTARY INFORMATION:

I. Background

    Upon request, FDA (the Agency or we) has classified the mechanical 
deviation device for esophageal protection during cardiac ablation 
procedures into class II (special controls), which we have determined 
will provide a reasonable assurance of the safety and effectiveness of 
the device. In addition, we believe this action will enhance patients' 
access to beneficial innovation, in part by reducing regulatory burdens 
by placing the device into a lower device class than the automatic 
class III assignment.
    The automatic assignment of class III occurs by operation of law 
and without any action by FDA, regardless of the level of risk posed by 
the new device. Any device that was not in commercial distribution 
before May 28, 1976, is automatically classified into, and remains 
within, class III and requires premarket approval unless and until FDA 
takes an action to classify or reclassify the device (21 U.S.C. 
360c(f)(1)). We refer to these devices as ``postamendments devices'' 
because they were not in commercial distribution prior to the date of 
enactment of the Medical Device Amendments of 1976, which amended the 
Federal Food, Drug, and Cosmetic Act (FD&C Act).
    FDA may take a variety of actions in appropriate circumstances to 
classify or reclassify a device into class I or II. We may issue an 
order finding a new device to be substantially equivalent under section 
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that 
does not require premarket approval. We determine whether a new device 
is substantially equivalent to a predicate device by means of the 
procedures for premarket notification under section 510(k) of the FD&C 
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
    FDA may also classify a device through ``De Novo'' classification, 
a common name for the process authorized under section 513(f)(2) of the 
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). 
Section 207 of the Food and Drug Administration Modernization Act of 
1997 (Pub. L. 105-115) established the first procedure for De Novo 
classification. Section 607 of the Food and Drug Administration Safety 
and Innovation Act (Pub. L. 112-144) modified the De Novo 
classification process by adding a second procedure. A device sponsor 
may utilize either procedure for De Novo classification.
    Under the first procedure, the person submits a premarket 
notification (510(k)) for a device that has not previously been 
classified. After receiving an order from FDA classifying the device 
into class III under section 513(f)(1) of the FD&C Act, the person then 
requests a classification under section 513(f)(2).
    Under the second procedure, rather than first submitting a 510(k) 
and then a request for classification, if the person determines that 
there is no legally marketed device upon which to base a determination 
of substantial equivalence, that person requests a classification under 
section 513(f)(2) of the FD&C Act.
    Under either procedure for De Novo classification, FDA is required 
to classify the device by written order within 120 days. The 
classification will be according to the criteria under section 
513(a)(1) of the FD&C Act. Although the device was automatically placed 
within class III, the De Novo classification is considered to be the 
initial classification of the device.
    We believe this De Novo classification will enhance patients' 
access to beneficial innovation, in part by reducing regulatory 
burdens. When FDA classifies a device into class I or II via the De 
Novo process, the device can serve as a predicate for future devices of 
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the 
FD&C Act). As a result, other device sponsors do not have to submit a 
De Novo request or premarket approval application to market a 
substantially equivalent device (see section 513(i) of the FD&C Act, 
defining ``substantial equivalence''). Instead, sponsors can use the 
less burdensome 510(k) process, when necessary, to market their device.

II. De Novo Classification

    On January 24, 2023, FDA received S4 Medical Corp.'s request for De 
Novo classification of the esolution Esophageal Retractor. FDA reviewed 
the request in order to classify the device under the criteria for 
classification set forth in section 513(a)(1) of the FD&C Act.
    We classify devices into class II if general controls by themselves 
are insufficient to provide reasonable assurance of the safety and 
effectiveness of the device, but there is sufficient information to 
establish special controls that, in combination with the general 
controls, provide reasonable assurance of the safety and effectiveness 
of the device for its intended use (see section 513(a)(1)(B) of the 
FD&C Act). After review of the information submitted in the request, we 
determined that the device can be classified into class II with the 
establishment of special controls. FDA has determined that these 
special controls, in addition to the general controls, will provide 
reasonable assurance of the safety and effectiveness of the device.
    Therefore, on September 6, 2023, FDA issued an order to the 
requester classifying the device into class II. In this final order, 
FDA is codifying the classification of the device by adding 21 CFR 
870.5710.\1\ We have named the generic type of device ``mechanical 
deviation device for esophageal protection during cardiac ablation 
procedures,'' and it is identified as a device that is placed in the 
lumen of the esophagus to reduce the likelihood of esophageal injury or 
a specific adverse event during cardiac ablation procedures. The device 
uses mechanical means to deviate the esophagus away from the source of 
ablation energy.
---------------------------------------------------------------------------

    \1\ FDA notes that the ``ACTION'' caption for this final order 
is styled as ``Final amendment; final order,'' rather than ``Final 
order.'' Beginning in December 2019, this editorial change was made 
to indicate that the document ``amends'' the Code of Federal 
Regulations. The change was made in accordance with the Office of 
Federal Register's (OFR) interpretations of the Federal Register Act 
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and 
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------

    FDA has identified the risks to health associated with this type of 
device and the measures required to mitigate these risks in table 1.

[[Page 57497]]



     Table 1--Risks to Health and Mitigation Measures for Mechanical
   Deviation Devices for Esophageal Protection During Cardiac Ablation
                               Procedures
------------------------------------------------------------------------
       Identified risks to health              Mitigation measures
------------------------------------------------------------------------
Failure to protect the esophagus during  Clinical performance testing;
 ablation leading to esophageal           animal performance testing;
 perforating complications.               non-clinical performance
                                          testing; and labeling.
Device malfunction leading to            Non-clinical performance
 esophageal injury.                       testing; and shelf life and
                                          packaging testing.
Adverse tissue reaction................  Biocompatibility evaluation.
Infection..............................  Sterilization validation; shelf
                                          life and packaging testing;
                                          and labeling.
Mechanical injury to esophageal or oral  Clinical performance testing;
 structures.                              animal performance testing;
                                          and labeling.
------------------------------------------------------------------------

    FDA has determined that special controls, in combination with the 
general controls, address these risks to health and provide reasonable 
assurance of the safety and effectiveness of the device. For a device 
to fall within this classification, and thus avoid automatic 
classification in class III, it would have to comply with the special 
controls named in this final order. The necessary special controls 
appear in the regulation codified by this final order. FDA supports the 
principles of the ``3Rs,'' to replace, reduce, and/or refine animal use 
in testing when feasible. We encourage sponsors to consult with us if 
they wish to use a non-animal testing method they believe is suitable, 
adequate, validated, and feasible. We will consider whether such an 
alternative method could be assessed for equivalency to an animal test 
method.
    Under the FD&C Act, submission of a premarket notification under 
section 510(k) is required to reasonably assure the safety and 
effectiveness of class II devices unless FDA determines that the device 
type should be exempt under section 510(m) of the FD&C Act. At this 
time FDA has not made this determination for mechanical deviation 
devices for esophageal protection during cardiac ablation procedures. 
This device is therefore subject to premarket notification requirements 
under section 510(k) of the FD&C Act.

III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is 
of a type that does not normally have a significant effect on the human 
environment. Therefore, neither an environmental assessment nor an 
environmental impact statement is required.

IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to 
previously approved collections of information found in other FDA 
regulations and guidance. These collections of information are subject 
to review by the Office of Management and Budget (OMB) under the 
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections 
of information in part 860, subpart D, regarding De Novo classification 
have been approved under OMB control number 0910-0844; the collections 
of information in 21 CFR part 814, subparts A through E, regarding 
premarket approval have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding 
premarket notification submissions have been approved under OMB control 
number 0910-0120; the collections of information in 21 CFR part 820 
regarding quality management system regulation have been approved under 
OMB control number 0910-0073; and the collections of information in 21 
CFR part 801 regarding labeling have been approved under OMB control 
number 0910-0485.

List of Subjects in 21 CFR Part 870

    Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
authority delegated to the Commissioner of Food and Drugs, 21 CFR part 
870 is amended as follows:

PART 870--CARDIOVASCULAR DEVICES

0
1. The authority citation for part 870 continues to read as follows:

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
2. Add Sec.  870.5710 to subpart F to read as follows:


Sec.  870.5710  Mechanical deviation device for esophageal protection 
during cardiac ablation procedures.

    (a) Identification. This device is placed in the lumen of the 
esophagus to reduce the likelihood of esophageal injury or a specific 
adverse event during cardiac ablation procedures. The device uses 
mechanical means to deviate the esophagus away from the source of 
ablation energy.
    (b) Classification. Class II (special controls). The special 
controls for this device are:
    (1) Clinical performance testing must demonstrate that the device 
performs as intended under anticipated conditions of use and include 
the following:
    (i) Evaluation of reduction of the incidence of esophageal injury 
during cardiac ablation procedures; and
    (ii) Evaluation of any esophageal or oral injury from use of the 
device.
    (2) Animal performance testing must demonstrate that the device 
performs as intended under the anticipated conditions of use and 
include the following:
    (i) Evaluation of the device's capability to adequately deviate the 
esophagus, including its trailing edge, away from the source of 
ablation energy; and
    (ii) Evaluation of any esophageal injury from use of the device.
    (3) Non-clinical performance testing must demonstrate that the 
device performs as intended under anticipated conditions of use and 
include the following:
    (i) Mechanical integrity testing using clinically relevant forces; 
and
    (ii) Compatibility testing with accessory devices.
    (4) Performance data must demonstrate the sterility of any device 
components intended to be provided sterile.
    (5) The patient-contacting components of the device must be 
demonstrated to be biocompatible.
    (6) Performance data must support the shelf life of the device by 
demonstrating package integrity and device functionality over the 
identified shelf life.
    (7) Labeling must include the following:
    (i) A summary of clinical performance testing with the device; and
    (ii) A shelf life.

Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-18433 Filed 9-9-26; 8:45 am]
BILLING CODE 4164-01-P