Medical Devices; Gastroenterology-Urology Devices; Classification of the Implanted Tibial Electrical Urinary Continence Device

Primary source

Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.

Published
September 10, 2026
Effective
September 10, 2026

Issuing agencies

Health and Human Services DepartmentFood and Drug Administration

Abstract

The Food and Drug Administration (FDA) is classifying the implanted tibial electrical urinary continence device into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for classification of the implanted tibial electrical urinary continence device. We are taking this action because we have determined that classifying the device into class II will provide a reasonable assurance of the safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

Full Text

<html>
<head>
<title>Federal Register, Volume 91 Issue 174 (Thursday, September 10, 2026)</title>
</head>
<body><pre>
[Federal Register Volume 91, Number 174 (Thursday, September 10, 2026)]
[Rules and Regulations]
[Pages 57504-57507]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2026-18428]


-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 876

[Docket No. FDA-2026-N-9908]


Medical Devices; Gastroenterology-Urology Devices; Classification 
of the Implanted Tibial Electrical Urinary Continence Device

AGENCY: Food and Drug Administration, HHS.

ACTION: Final amendment; final order.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA) is classifying the 
implanted tibial electrical urinary continence device into class II 
(special controls). The special controls that apply to the device type 
are identified in this order and will be part of the codified language 
for classification of the implanted tibial electrical urinary 
continence device. We are taking this action because we have determined 
that classifying the device into class II will provide a reasonable 
assurance of the safety and effectiveness of the device. We believe 
this action will also enhance patients' access to beneficial innovative 
devices, in part by reducing regulatory burdens.

DATES: This order is effective September 10, 2026. The classification 
was applicable on August 16, 2023.

FOR FURTHER INFORMATION CONTACT: Sharon Andrews, Center for Devices and 
Radiological Health, Food and Drug Administration, 10903 New Hampshire 
Ave., Bldg. 66, Rm. 2640, Silver Spring, MD 20993-0002, 301-796-6529, 
<a href="/cdn-cgi/l/email-protection#d685beb7a4b9b8f897b8b2a4b3a1a596b0b2b7f8bebea5f8b1b9a0"><span class="__cf_email__" data-cfemail="d083b8b1a2bfbefe91beb4a2b5a7a390b6b4b1feb8b8a3feb7bfa6">[email&#160;protected]</span></a>.

SUPPLEMENTARY INFORMATION:

I. Background

    Upon request, FDA (the Agency or we) has classified the implanted 
tibial electrical urinary continence device into class II (special 
controls), which we have determined will provide a reasonable assurance 
of the safety and effectiveness of the device. In addition, we believe 
this action will enhance

[[Page 57505]]

patients' access to beneficial innovation, in part by reducing 
regulatory burdens by placing the device into a lower device class than 
the automatic class III assignment.
    The automatic assignment of class III occurs by operation of law 
and without any action by FDA, regardless of the level of risk posed by 
the new device. Any device that was not in commercial distribution 
before May 28, 1976, is automatically classified into, and remains 
within, class III and requires premarket approval unless and until FDA 
takes an action to classify or reclassify the device (21 U.S.C. 
360c(f)(1)). We refer to these devices as ``postamendments devices'' 
because they were not in commercial distribution prior to the date of 
enactment of the Medical Device Amendments of 1976, which amended the 
Federal Food, Drug, and Cosmetic Act (FD&C Act).
    FDA may take a variety of actions in appropriate circumstances to 
classify or reclassify a device into class I or II. We may issue an 
order finding a new device to be substantially equivalent under section 
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that 
does not require premarket approval. We determine whether a new device 
is substantially equivalent to a predicate device by means of the 
procedures for premarket notification under section 510(k) of the FD&C 
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
    FDA may also classify a device through ``De Novo'' classification, 
a common name for the process authorized under section 513(f)(2) of the 
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). 
Section 207 of the Food and Drug Administration Modernization Act of 
1997 (Pub. L. 105-115) established the first procedure for De Novo 
classification. Section 607 of the Food and Drug Administration Safety 
and Innovation Act (Pub. L. 112-144) modified the De Novo 
classification process by adding a second procedure. A device sponsor 
may utilize either procedure for De Novo classification.
    Under the first procedure, the person submits a premarket 
notification (510(k)) for a device that has not previously been 
classified. After receiving an order from FDA classifying the device 
into class III under section 513(f)(1) of the FD&C Act, the person then 
requests a classification under section 513(f)(2).
    Under the second procedure, rather than first submitting a 510(k) 
and then a request for classification, if the person determines that 
there is no legally marketed device upon which to base a determination 
of substantial equivalence, that person requests a classification under 
section 513(f)(2) of the FD&C Act.
    Under either procedure for De Novo classification, FDA is required 
to classify the device by written order within 120 days. The 
classification will be according to the criteria under section 
513(a)(1) of the FD&C Act. Although the device was automatically placed 
within class III, the De Novo classification is considered to be the 
initial classification of the device.
    We believe this De Novo classification will enhance patients' 
access to beneficial innovation, in part by reducing regulatory 
burdens. When FDA classifies a device into class I or II via the De 
Novo process, the device can serve as a predicate for future devices of 
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the 
FD&C Act). As a result, other device sponsors do not have to submit a 
De Novo request or premarket approval application to market a 
substantially equivalent device (see section 513(i) of the FD&C Act, 
defining ``substantial equivalence''). Instead, sponsors can use the 
less burdensome 510(k) process, when necessary, to market their device.

II. De Novo Classification

    On October 5, 2022, FDA received BlueWind Medical Ltd.'s request 
for De Novo classification of the Revi System. FDA reviewed the request 
in order to classify the device under the criteria for classification 
set forth in section 513(a)(1) of the FD&C Act.
    We classify devices into class II if general controls by themselves 
are insufficient to provide reasonable assurance of the safety and 
effectiveness of the device, but there is sufficient information to 
establish special controls that, in combination with the general 
controls, provide reasonable assurance of the safety and effectiveness 
of the device for its intended use (see section 513(a)(1)(B) of the 
FD&C Act). After review of the information submitted in the request, we 
determined that the device can be classified into class II with the 
establishment of special controls. FDA has determined that these 
special controls, in addition to the general controls, will provide 
reasonable assurance of the safety and effectiveness of the device.
    Therefore, on August 16, 2023, FDA issued an order to the requester 
classifying the device into class II. In this final order, FDA is 
codifying the classification of the device by adding 21 CFR 
876.5305.\1\ We have named the generic type of device ``implanted 
tibial electrical urinary continence device,'' and it is identified as 
an implanted prescription device that receives power from a non-
implanted external power source to provide electrical stimulation of 
the tibial nerve in proximity to the ankle. The device is intended for 
the treatment of overactive bladder related symptoms of urge urinary 
incontinence, urinary urgency, urinary frequency, and nocturia.
---------------------------------------------------------------------------

    \1\ FDA notes that the ``ACTION'' caption for this final order 
is styled as ``Final amendment; final order,'' rather than ``Final 
order.'' Beginning in December 2019, this editorial change was made 
to indicate that the document ``amends'' the Code of Federal 
Regulations. The change was made in accordance with the Office of 
Federal Register's (OFR) interpretations of the Federal Register Act 
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and 
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------

    FDA has identified the risks to health associated with this type of 
device and the measures required to mitigate these risks in table 1.

  Table 1--Risks to Health and Mitigation Measures for Implanted Tibial
                  Electrical Urinary Continence Devices
------------------------------------------------------------------------
       Identified risks to health              Mitigation measures
------------------------------------------------------------------------
Overstimulation leading to nerve/tissue  Non-clinical performance
 damage.                                  testing; Electromagnetic
                                          compatibility testing;
                                          Electrical safety testing;
                                          Software verification,
                                          validation, and hazard
                                          analysis; Wireless coexistence
                                          testing; and Labeling.
Adverse tissue reaction................  Biocompatibility evaluation;
                                          and Labeling.
Infection..............................  Sterilization validation; Shelf-
                                          life testing; and Labeling.

[[Page 57506]]

 
Thermal injury.........................  Non-clinical performance
                                          testing; Thermal safety
                                          testing; and Electrical safety
                                          testing.
Interference with other medical devices  Electromagnetic compatibility
                                          testing; Magnetic resonance
                                          compatibility testing;
                                          Electrical safety testing;
                                          Software verification,
                                          validation and hazard
                                          analysis; Wireless coexistence
                                          testing; and Labeling.
Pain and discomfort....................  Non-clinical performance
                                          testing; Electrical safety
                                          testing; and Labeling.
Electrical shock or stimulation of non-  Electrical safety testing;
 target tissue.                           Electromagnetic compatibility
                                          testing; and Labeling.
Mechanical injury to device or tissue/   Non-clinical performance
 nerves.                                  testing; and Labeling.
Undesired fluid retention due to use in  Labeling.
 inappropriate population.
------------------------------------------------------------------------

    FDA has determined that special controls, in combination with the 
general controls, address these risks to health and provide reasonable 
assurance of the safety and effectiveness of the device. For a device 
to fall within this classification, and thus avoid automatic 
classification in class III, it would have to comply with the special 
controls named in this final order. The necessary special controls 
appear in the regulation codified by this final order.
    At the time of classification, implanted tibial electrical urinary 
continence devices are for prescription use only. Prescription devices 
are exempt from the requirement for adequate directions for use for the 
layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) 
and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met.
    Under the FD&C Act, submission of a premarket notification under 
section 510(k) is required to reasonably assure the safety and 
effectiveness of class II devices unless FDA determines that the device 
type should be exempt under section 510(m) of the FD&C Act. At this 
time FDA has not made this determination for implanted tibial 
electrical urinary continence devices. This device is therefore subject 
to premarket notification requirements under section 510(k) of the FD&C 
Act.

III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is 
of a type that does not normally have a significant effect on the human 
environment. Therefore, neither an environmental assessment nor an 
environmental impact statement is required.

IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to 
previously approved collections of information found in other FDA 
regulations and guidance. These collections of information are subject 
to review by the Office of Management and Budget (OMB) under the 
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections 
of information in part 860, subpart D, regarding De Novo classification 
have been approved under OMB control number 0910-0844; the collections 
of information in 21 CFR part 814, subparts A through E, regarding 
premarket approval have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding 
premarket notification submissions have been approved under OMB control 
number 0910-0120; the collections of information in 21 CFR part 820 
regarding quality management system regulation have been approved under 
OMB control number 0910-0073; and the collections of information in 21 
CFR part 801 regarding labeling have been approved under OMB control 
number 0910-0485.

List of Subjects in 21 CFR Part 876

    Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
authority delegated to the Commissioner of Food and Drugs, 21 CFR part 
876 is amended as follows:

PART 876--GASTROENTEROLOGY-UROLOGY DEVICES

0
1. The authority citation for part 876 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
2. Add Sec.  876.5305 to subpart F to read as follows:


Sec.  876.5305  Implanted tibial electrical urinary continence device.

    (a) Identification. An implanted tibial electrical urinary 
continence device is an implanted prescription device that receives 
power from a non-implanted external power source to provide electrical 
stimulation of the tibial nerve in proximity to the ankle. The device 
is intended for the treatment of overactive bladder related symptoms of 
urge urinary incontinence, urinary urgency, urinary frequency, and 
nocturia.
    (b) Classification. Class II (special controls). The special 
controls for this device are:
    (1) Non-clinical performance testing must demonstrate that the 
device performs as intended under anticipated conditions of use. The 
following testing must be conducted:
    (i) Electrical performance testing of the device must be conducted 
to validate the specified electrical output and duration of stimulation 
of the device; and
    (ii) Testing must verify the implant can withstand clinically 
relevant forces during and after implantation.
    (2) The patient-contacting components of the device must be 
demonstrated to be biocompatible.
    (3) Performance data must demonstrate the sterility of the patient-
contacting components of the device.
    (4) Performance data must support the shelf life of the device by 
demonstrating continued sterility of patient-contacting components, 
package integrity, and device functionality over the identified shelf 
life.
    (5) Performance testing must demonstrate the electromagnetic 
compatibility, electrical safety, thermal safety, and wireless 
performance of the device.
    (6) Software verification, validation, and hazard analysis must be 
performed.
    (7) Performance testing must evaluate the compatibility of the 
device in a magnetic resonance environment.
    (8) Labeling for the device must include:
    (i) A contraindication against use during pregnancy;
    (ii) A contraindication against using the device in men who have 
Benign Prostatic Hyperplasia (BPH) or other lower urinary tract 
obstructions;

[[Page 57507]]

    (iii) A detailed summary of the device technical parameters and the 
typical course of treatment;
    (iv) Device- and procedure-related adverse events pertinent to use 
of the device; and
    (v) A shelf life for any sterile components.
    (9) Patient labeling must include:
    (i) Post-operative care instructions to avoid infection and 
inflammation of the surgical site;
    (ii) Instructions to avoid overstimulation related nerve and tissue 
damage;
    (iii) Instructions to avoid mechanical injuries to nerve and tissue 
caused by the implanted component;
    (iv) Instructions for reprocessing/cleaning of any reusable 
components;
    (v) Clinical performance reported by relevant subgroups;
    (vi) The risks and benefits associated with use of the device;
    (vii) Information on the typical course of treatment;
    (viii) Instructions to avoid pain and discomfort; and
    (ix) Instructions to avoid interference with other medical devices.

Grace R. Graham,
Deputy Commissioner for Policy, Legislation, and International Affairs.
[FR Doc. 2026-18428 Filed 9-9-26; 8:45 am]
BILLING CODE 4164-01-P