Medical Devices; General Hospital and Personal Use Devices; Classification of the Intravenous Catheter Force-Activated Separation Device
Primary source
Metadata and text below are from the Federal Register, a public-domain U.S. government work. Always verify the official published version before relying on it for any legal matter.
Issuing agencies
Abstract
The Food and Drug Administration (FDA, Agency, or we) is classifying the intravenous catheter force-activated separation device into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the intravenous catheter force-activated separation device's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.
Full Text
<html>
<head>
<title>Federal Register, Volume 89 Issue 159 (Friday, August 16, 2024)</title>
</head>
<body><pre>
[Federal Register Volume 89, Number 159 (Friday, August 16, 2024)]
[Rules and Regulations]
[Pages 66558-66560]
From the Federal Register Online via the Government Publishing Office [<a href="http://www.gpo.gov">www.gpo.gov</a>]
[FR Doc No: 2024-18267]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA-2024-N-3356]
Medical Devices; General Hospital and Personal Use Devices;
Classification of the Intravenous Catheter Force-Activated Separation
Device
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, Agency, or we) is
classifying the intravenous catheter force-activated separation device
into class II (special controls). The special controls that apply to
the device type are identified in this order and will be part of the
codified language for the intravenous catheter force-activated
separation device's classification. We are taking this action because
we have determined that classifying the device into class II (special
controls) will provide a reasonable assurance of safety and
effectiveness of the device. We believe this action will also enhance
patients' access to beneficial innovative devices.
DATES: This order is effective August 16, 2024. The classification was
applicable on May 27, 2021.
FOR FURTHER INFORMATION CONTACT: Florencia Wilson, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 2458, Silver Spring,
[[Page 66559]]
MD 20993-0002, 240-402-9978, <a href="/cdn-cgi/l/email-protection#5f1933302d3a313c363e710836332c30311f393b3e7137372c71383029"><span class="__cf_email__" data-cfemail="c086acafb2a5aea3a9a1ee97a9acb3afae80a6a4a1eea8a8b3eea7afb6">[email protected]</span></a>.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the intravenous catheter force-
activated separation device as class II (special controls), which we
have determined will provide a reasonable assurance of safety and
effectiveness. In addition, we believe this action will enhance
patients' access to beneficial innovation, in part by placing the
device into a lower device class than the automatic class III
assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device
that does not require premarket approval. We determine whether a new
device is substantially equivalent to a predicate device by means of
the procedures for premarket notification under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 (Pub. L. 105-115) established the first procedure for De
Novo classification. Section 607 of the Food and Drug Administration
Safety and Innovation Act (Pub. L. 112-144) modified the De Novo
application process by adding a second procedure. A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
When FDA classifies a device into class I or II via the De Novo
process, the device can serve as a predicate for future devices of that
type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C
Act). As a result, other device sponsors do not have to submit a De
Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less-burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On September 18, 2019, FDA received Site Saver, Inc. d/b/a Lineus
Medical's request for De Novo classification of the SafeBreak Vascular.
FDA reviewed the request to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on May 27, 2021, FDA issued an order to the requester
classifying the device into class II. In this final order, FDA is
codifying the classification of the device by adding 21 CFR
880.5220.\1\ We have named the generic type of device intravenous
catheter force-activated separation device and it is identified as a
device placed in-line with an intravenous (IV) catheter and an
intravascular administration set, including any administration set
accessories. It separates into two parts when a specified force is
applied. The device is intended to reduce the risk of IV catheter
failure(s) requiring IV catheter replacement.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Intravenous Catheter Force-Activated Separation Device Risks
and Mitigation Measures
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Delays of therapy due to failure of Performance data, Non-clinical
device to function as expected (e.g., performance testing, and
if separation force too low). Labeling.
Mechanical complications (e.g., IV Performance data, Non-clinical
dislodgement, IV infiltration, performance testing, and
occlusion, and phlebitis events Labeling.
requiring IV replacement) due to
failure of device to function as
expected (e.g., if separation force
too high).
Infection.............................. Sterilization validation, Shelf
life testing, Non-clinical
performance testing, and
Labeling.
Air embolism........................... Non-clinical performance
testing, and Labeling.
[[Page 66560]]
Adverse tissue reaction................ Biocompatibility evaluation,
Pyrogenicity testing, and Non-
clinical performance testing.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in 21 CFR part 860, subpart D, regarding De Novo
classification have been approved under OMB control number 0910-0844;
the collections of information in 21 CFR part 814, subparts A through
E, regarding premarket approval, have been approved under OMB control
number 0910-0231; the collections of information in part 807, subpart
E, regarding premarket notification submissions, have been approved
under OMB control number 0910-0120; the collections of information in
21 CFR part 820, regarding quality system regulation, have been
approved under OMB control number 0910-0073; and the collections of
information in 21 CFR part 801, regarding labeling, have been approved
under OMB control number 0910-0485.
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
880 is amended as follows:
PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES
0
1. The authority citation for part 880 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 880.5220 to read as follows:
Sec. 880.5220 Intravenous catheter force-activated separation device.
(a) Identification. An intravenous catheter force-activated
separation device is placed in-line with an intravenous (IV) catheter
and an intravascular administration set, including any administration
set accessories. It separates into two parts when a specified force is
applied. The device is intended to reduce the risk of IV catheter
failure(s) requiring IV catheter replacement.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Performance data must be provided to demonstrate clinically
acceptable performance for the intended use of the device.
(2) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance characteristics must be tested:
(i) Separation force testing;
(ii) Validation of anti-reconnect features;
(iii) Air and liquid leakage testing, both before and after
separation;
(iv) Luer connection testing;
(v) Flow rate testing;
(vi) Particulate testing; and
(vii) Microbial ingress testing.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate that the device is sterile
and non-pyrogenic.
(5) Performance testing must support the shelf life of the device
by demonstrating continued sterility and device functionality over the
identified shelf life.
(6) Device labeling must include:
(i) Instructions for use; and
(ii) A discussion of catheter dressings intended to be used with
the device.
Dated: August 12, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024-18267 Filed 8-15-24; 8:45 am]
BILLING CODE 4164-01-P